(151)
Nitrogen-total LR
(all steps)
0.5 - 25.0
mg/l (N)
Reagent-Kit
Internal Name: 151-NitroTotLR-pre
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
Nitrogen LR Kit (PL151-Kit)
Measurement procedure:
1
Prepare 2 Hydroxide LR cuvettes (16 mm). Label one as a ZERO cuvette.
2
Add 1 x Persulfate powder pillow into both cuvettes.
3
Open the first cuvette (ZERO cuvette) and add 2 ml distilled water.
4
Open the second cuvette (sample cuvette) and add 2 ml sample water.
5
Screw the lid immediately back onto both cuvettes.
6
Shake the vials vigorously for 00:30 minute(s).
7
Tap TEST to start a 00:30 minute(s) countdown.
8
Place cuvettes for 30 minutes at 100 °C in the preheated thermoreactor.
9
Tap TEST to start a 30:00 minute(s) countdown.
10
CAUTION: Cuvettes are hot!
11
Remove the cuvettes from the thermoreactor.
12
Let the cuvettes cool down to at least 25 °C.
13
Add 1 x Reagent A powder pillow into both cuvettes.
14
Screw the lid immediately back onto both cuvettes.
15
Shake the vials vigorously for 00:20 minute(s).
16
Tap TEST to start a 00:20 minute(s) countdown.
17
The reagents should now react.
18
Tap TEST to start a 03:00 minute(s) countdown.
19
Add 1 x Reagent B powder pillow into both cuvettes.
20
Screw the lid back on both cuvettes.
21
Shake the vials vigorously for 00:20 minute(s).
22
Tap TEST to start a 00:20 minute(s) countdown.
23
The reagents should now react.
24
Tap TEST to start a 02:00 minute(s) countdown.
25
Open 1 Acid LR/HR cuvette and add 2 ml sample water from the previous ZERO cuvette.
This is your new ZERO cuvette.
26
Add 2 ml from the previous test cuvette into a new cuvette. This is your new test cuvette.
T133/204
T133/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...