(27)
Iodine
0.00 - 28.00
mg/l (I₂)
Tablet
Internal Name: 27-Iodine-tab
10ml
DPD N°1 Photometer (TbsPD1)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Empty the cuvette to a few drops.
8
Add 1 DPD N°1 Photometer tablet(s) to the test water in the cuvette.
9
Crush the tablet with a clean stirring rod.
10
Fill the cuvette to 10 ml with the sample water.
11
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid well.
14
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
15
Tap TEST to start a 00:10 minute(s) countdown.
16
After the lapse of a 00:10 minute(s) countdown the measured value is displayed.
Notes:
Water samples with a high calcium content resp. a high conductivity will render the sample
cloudy, which is detrimental to the measurement precision. In this case use "DPD N°1 High
Calcium (HC)" reagents.
Do not clean glassware and equipment with household detergents, as these could greatly
reduce the measurement result. To prevent any contamination, the cuvette, lid and stirrer
should be stored in a 0.1 % sodium hypochlorite solution for one hour. Then rinse with distilled
water, thoroughly.
T96/204
T96/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...