27
Open 2 Acid LR/HR cuvette and add 2 ml sample water from the previous ZERO cuvette.
This is your new ZERO cuvette.
28
Add 2 ml from the previous test cuvette into a new cuvette. This is your new test cuvette.
29
Screw the lid back on both cuvettes.
30
Gently swirl both cuvettes to mix the liquids well. Caution, heat development!
31
Place the 16 mm adapter in the PrimeLab.
32
Place the ZERO cuvette in the PrimeLab.
33
Put on the lightshield.
34
Start ZERO measurement.
35
Remove the cuvette again.
36
Insert the sample cuvette in the PrimeLab.
37
Put on the lightshield.
38
Tap TEST to perform the measurement.
Notes:
The test result can be converted into the following unit(s): mg/l NH₃, mg/l NH₄⁺.
This test can be used for water, wastewater and seawater.
Remove powder from vial edges, lid and tube threads after adding powder.
Use volumetric pipettes to dose exactly 2 ml of the Acid LR/HR reagent.
Reagents might not dissolve entirely.
Incubation time shall NOT exceed 30 minutes!
Step 30 should be performed by turning vial upside down and back. Waiting for the solution to
entirely flow back and inverse 10 times.
Zero vial can be stored and used for max. 7 days if stored in the dark
Dilute and repeat measurement if large quantities of nitrogen compounds (free, organic) are
present, as they may interfere and reduce the effectiveness of the digestion.
Bromide concentrations > 60 mg/l and chloride concentrations > 1000 mg/l interfere and lead to
results increased by 10 %.
T137/204
T137/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...