TRACER
0 - 1000
µg/l (TraceR)
Internal Name: 157-TraceR-Ad
10ml
Prime L ab
Measurement procedure:
1
Fill 10 ml distilled water in a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Empty the cuvette.
8
Rinse the cuvette with the test water.
9
Fill 10 ml test water into a clean 24 mm cuvette.
10
Screw the lid back on the cuvette.
11
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
12
Tap TEST to perform the measurement.
13
The measured value is immediately displayed.
Notes:
It is essential to always ensure the correct amount of water in the cell, which is why exactly 10
ml of liquid should be taken by the pipette for the subsequent sample measurement. Please
change or clean the tip of the pipette after each measurement/calibration.
T176/204
T176/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...