(45)
Phosphate
(-ortho-) LR
0.00 - 4.00
mg/l (PO₄³⁻)
Liquid
+ Powder
Internal Name: 45-Phosphat-LR-liq
10ml
65ml PL Phosphate LR N°1 (PL65PPLR1)
PL Phosphate LR 2 (PLpow20PPLR2)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 25 drop(s) of PL Phosphate LR 1 into the cuvette.
8
Add 1 x 0.05 ml PL Phosphate LR 2 scoop(s) powder to the sample water in the cuvette.
9
Screw the lid back on the cuvette.
10
Gently swirl the cuvette to mix the liquid well.
11
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
12
Tap TEST to perform the measurement.
13
After the lapse of a 10:00 minute(s) countdown the measured value is displayed.
Notes:
The test result can be converted into the following unit(s): mg/l P, mg/l P₂O₅.
With this procedure ortho-phosphate ions are detected. Other phosphates/phosphonates must
therefore be converted into ortho-phosphates before the test is begun.
The pH value of the sample water should be between 6 and 7.
The following contents of substances in the sample water can - at the respective concentration -
falsify the measurement results: Chromium (> 100 mg/l), copper (> 10 mg/l), iron (> 100 mg/l),
nickel (> 300 mg/l), zinc (> 80 mg/l).
T159/204
T159/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...