(159)
Permanganate Time Test
0 - 100
%T (PTT)
Liquid
Internal Name: 159-PTT-tab
10ml
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
Potassium Permanganate Solution
(PL10PTT)
Measurement procedure:
1
Set the refrigerator to 15 °C using the thermostat and according to the instructions.
2
Fill 10 ml test water into a clean 24 mm cuvette.
3
Screw the lid back on the cuvette.
4
Place the cuvette in the refrigerator for 20 minute(s).
5
Place the sealed "Methanol ZERO" cuvette in the PrimeLab.
6
Start ZERO measurement.
7
Remove the cuvette from the PrimeLab and set it aside. It is no longer required for this test.
8
Remove the cuvette from the refrigerator.
9
Add exactly 35 µl of Potassium Permanganate Solution to the sample water into the same
cuvette. Use a 10-100 µl pipette.
10
Screw the lid back on the cuvette.
11
Place the cuvette in the refrigerator for 10 minute(s).
12
Tap TEST to start a 10:00 minute(s) countdown.
13
Remove the cuvette from the refrigerator.
14
Wipe the condensation water from the cuvette with a dry cloth.
15
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
16
Tap TEST to perform the measurement.
Notes:
After adding the PTT-liquid, sample is highly sensitive to light, air and temperature. Do NOT
open the vial after PTT-liquid has been added and lid got screwed on and keep it at constant
temperature of 15 °C.
Interferences: Turbid and/or colored water samples (before adding PTT liquid).
T150/204
T150/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...