45
Super OTZ
PrimeLab
2.0
Tap on "Super OTZ" in the main menu to display all OTZs that already exist.
• Follow the on-screen instructions (to go to the next step tap the single arrow icon , to go directly
to the ZERO measurement tap the double arrow icon. )
• Tap the hamburger menu in the top right corner.
• Name your OTZ anything you want to identify it later.
• Tap on "Add OTZ".
• Tap "Start Super OTZ" to begin the ZERO measurement.
• Select the parameter to which you want to assign the OTZ.
Add Super OTZ
You have several options to delete an OTZ.
• Tap and hold on the OTZ until it turns blue. By tapping on more OTZs, you can add them
to the selection.
• Swipe the OTZ you want to delete to the left, and then tap the trash can icon.
// Alternatively, tap on the hamburger menu and then on "Delete selected OTZ".
Tap on "Super OTZ" in the main menu to display all OTZs that already exist.
• Tap on "Delete selected" at the bottom left to delete the selected OTZs.
Delete Super OTZ
Search Super OTZ
Tap on "Super OTZ" in the main menu to display all OTZs that already exist.
You have several options to search for an OTZ.
• Tap on the magnifying glass icon in the upper right corner and enter the desired search term in
the search field.
• You can search for whole phrases or just fractions of them.
// Alternatively, tap on the hamburger menu and then on "Search" . Then enter the desired
search term in the search field.
To ensure a faster measurement process for repetitive measuring of sapmles (e.g. COD), there is
now the option of Super OTZ (One Time ZERO).
With this function, a ZERO measurement is stored in the PrimeLab 2.0 and can be recalled for each
measurement.
The ZERO value determines a color/turbidity of your water sample in order to eliminate any
preliminary coloration or turbidity. PrimeLab 2.0 stores the last ZERO value to be able to perform
more than one test with the same undiluted water source (!) without having to repeat the ZERO
measurement each time. Since some parameters use different ZERO methods, such as 10ml water
sample or 5ml water sample plus 5ml deionized water, PrimeLab 2.0 recognizes the ZERO type of
each measurement and offers OTZ only if the following measurement matches the ZERO type of the
last performed measurement.
Almost every test requires a ZERO measurement.
Intelligent OTZ (One-Time-Zero)
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...