Phosphate
(-ortho-) HR
0.0 - 100.0
mg/l (PO₄³⁻)
Liquid
Internal Name: 47-Phosphat-HR-liq
10ml
65ml PL Phosphate HR N°1
(PL65PPHR1)
65ml PL Phosphate HR N°2
(PL65PPHR2)
Measurement procedure:
1
Separate the two halves of the filter holder.
2
Insert a 25 mm (GF/C) filter. Screw the filter holder back together, making sure that the O-ring
is correctly seated.
3
Fill a clean 20 ml syringe with 14 ml sample water.
4
Connect the syringe to the filter holder.
5
Empty the syringe with the filter up to the 10 ml mark.
6
Fill the remaining 10 ml filtered sample water into a clean 24 mm cuvette.
7
Screw the lid back on the cuvette.
8
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
9
Start ZERO measurement.
10
Remove the cuvette again.
11
Unscrew the lid from the cuvette.
12
Add 25 (~ 1 ml) drop(s) of PL Phosphate HR 1 into the cuvette.
13
Add 25 (~ 1 ml) drop(s) of PL Phosphate HR 2 into the cuvette.
14
Screw the lid back on the cuvette.
15
Gently swirl the cuvette to mix the liquid well.
16
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
17
Tap TEST to perform the measurement.
18
After the lapse of a 10:00 minute(s) countdown the measured value is displayed.
Notes:
Special accessories required / not included as standard equipment!
The test result can be converted into the following unit(s): mg/l P, mg/l P₂O₅.
Filter process is only needed in case of any suspended insoluble phosphate expected in your
water sample (applicable for boiler water testing)
With this procedure ortho-phosphate ions are detected. Other phosphates/phosphonates must
therefore be converted into ortho-phosphates before the test is begun.
T162/204
T162/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...