
(130)
Chlorine Dioxide
0.00 - 5.00
mg/l (ClO₂)
Liquid
+ Powder Pack
Internal Name: 130-Chl-Diox-pp
10ml
30ml PL DPD Glycine (PL30DPDGlycine)
DPD N° 1 Photometer (PPPD150)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 4 drop(s) of PL DPD Glycine into the cuvette.
8
Screw the lid back on the cuvette.
9
Gently swirl the cuvette to mix the liquid well.
10
Add 1 DPD N°1 Photometer powder pillow to the sample water in the cuvette.
11
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid well.
14
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
15
Tap TEST to start a 02:00 minute(s) countdown.
16
After the lapse of a 02:00 minute(s) countdown the measured value is displayed.
Notes:
If the sample water contains more than 30 mg/l chlorine dioxide, a measured value of 0 mg/l
can be displayed. In that case, a dilution is recommended.
If the sample water contains further oxidizing agents, these will react like chlorine dioxide and
contribute to the measurement result.
Make sure no chlorine dioxide escapes while preparing the sample. The measurement must be
performed directly after sampling.
The DPD reagent buffers the pH value of the sample water in the range of 6.2 to 6.5, which is
essential for colour development. If the sample water is very alkaline or acidic, this must be
adjusted to a pH within the range of 6 to 7 by addition of 0.5 mol/l sulphuric acid or 1 mol/l
caustic soda, respectively, before the DPD reagent is added.
T47/204
T47/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...