(171)
Iron in Oil
20 - 450
mg/l (Fe²⁺)
Internal Name: 171-IronInOil-tab
Measurement procedure:
1
Take a semi-micro vial cuvette and label it as "ZERO".
2
Transfer approx. 1 ml of Reagent S into the cuvette.
3
Place the ZERO cuvette into the cuvette holder of the photometer.
4
Put on the light shield and press "ZERO".
5
Remove the light shield.
6
Remove the cuvette again.
7
SAMPLE TESTING: Label the reaction vials and cuvettes according to the number of
samples to be tested (e.g. 1..2..3..).
8
Transfer 10 ml of Reagent W into each reaction vial.
9
Transfer approx. 3 ml of Reagent S into each reaction vial by using a Pasteur pipette.
10
Transfer into each reaction vial individually 0.1 ml of the well mixed cylinder drain oil sample.
11
Transfer 1 Iron-reaction-tablet into each reaction vial.
12
Close the reaction vials.
13
Tap "NEXT" to start a 02:00 minute(s) shake coutdown and shake all reaction vials
immediately.
14
Tap "NEXT" to start a 10:00 minute(s) reaction countdown.
15
Shake again all reaction vials. Press "NEXT" to start a 15 second(s) countdown.
16
Wait for the phase separation in the reaction vials.
17
Take about 3.5 ml of the lower magenta (light to strong) coloured phase with a clean 5 ml
syringe.
18
At the tip of this syringe, connect a 0.45 µm syringe filter.
19
Filter the magenta solution into a clean cuvette.
20
Insert the cuvette into the cuvette holder of the photometer.
21
Cover with the light shield and press "TEST". Note the Iron value of the sample in mg/l.
T139/204
T139/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...