(38)
pH-value MR
6.50 - 8.40
(pH)
Tablet
Internal Name: 38-pH-MR-tab
10ml
Phenol Red Photometer (TbsPph)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 1 Phenol Red Photometer tablet(s) to the test water in the cuvette.
8
Crush the tablet with a clean stirring rod.
9
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
10
Screw the lid back on the cuvette.
11
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
12
Tap TEST to perform the measurement.
13
The measured value is immediately displayed.
Notes:
PH value out of the range of pH 6.5 to 8.4 can lead to incorrect test results. If you are not sure,
a control measurement by pH electrode is recommended.
An alkalinity level > 40 mg/l CaCO₃ is required to prevent inaccurate readings.
Depending on the salt content of the water sample, the measurement result must be manually
corrected according to the following scheme: 1 molar = -0.21 pH; 2 molar = -0.26 pH; 3 molar =
-0.29 pH with: 1 mol of salt (NaCl) = 5.8 % = 58.4 g/l.
T151/204
T151/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...