Nitrite LR
0.00 - 0.50
mg/l (N)
Powder Pack
Internal Name: 35-Nitrite-LR-tab
10ml
Nitrite LR Photometer (PPPNiLR)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 1 Nitrite LR Photometer powder pillow to the sample water in the cuvette.
8
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
9
Screw the lid back on the cuvette.
10
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
11
Tap TEST to start a 10:00 minute(s) countdown.
12
After the lapse of a 10:00 minute(s) countdown the measured value is displayed.
Notes:
The test result can be converted into the following unit(s): mg/l NaNO₂, mg/l NO₂⁻.
T128/204
T128/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...