(70)
Erythorbic Acid
0.00 - 3.50
mg/l (EA)
Liquid
Internal Name: 70-Erythorbic-Acid
10ml
PL Oxygen Scavenger 1 (65 ml) (PL65OxyScav1)
PL Oxygen Scavenger 2 (65 ml) (PL65OxyScav2)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 6 drop(s) of PL Oxygen Scavenger 1 into the cuvette.
8
Add 25 drop(s) of PL Oxygen Scavenger 2 into the cuvette.
9
Screw the lid back on the cuvette.
10
Gently swirl the cuvette to mix the liquid well.
11
Tap TEST to perform the measurement.
12
After the lapse of a 10:00 minute(s) countdown the measured value is displayed.