Avoid contact with eyes. Wear protective gloves.
Certain isolates cannot be detected below 106 cfu.
Disposal of product according to local regulations. Products are stable and unlikely to react in a
hazardous manner under normal conditions of use.
Do NOT re-use semi-micro vials (LG-CB).
Leave at least 12 cm space between multiple LG-MH (magnetic holders).
Reagents are supplied in excess. Do NOT re-use any leftover amounts of reagents.
When emptying cuvettes LG-MHCB, always do so to the BACK and never in front (magnet)!
Please follow the test procedure properly to avoid misreadings.
Once lids of LG-MHCB are removed and to discarded, do NOT use them for any of the following
test steps.
If you do more than 1 test at the same time, only one blanc/ZERO vial is needed.
We propose to use LG-MP4 automatic agitator plate to place up to 20 LG-MHCB cuvettes in
case you do multiple tests at one time.
Measurement has to be performed immediately after the last step (countdown), as the color
reaction might continue.
Leaving reagents at room temperature for 30 minutes before starting the test is essential.
When using larger units of reagents, immediately restore in fridge after use.
Depending on the water quality of the test water, the pre-filter have to be changed during the
filter process, if it is too dirty.
T112/204
T112/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...