(107)
Colour
(true)
15 - 500
mg/l (Pt-Co)
Internal Name: 107-Colour (Hazen/APHA)
10ml
Measurement procedure:
1
Separate the two halves of the filter holder.
2
Insert a 0.45 µm filter. Screw the filter holder back together, making sure that the O-ring is
correctly seated.
3
Fill the syringe again with distilled water.
4
Connect the syringe to the filter holder.
5
Empty the syringe with the filter completely.
6
Remove the filter syringe from the filter holder.
7
Repeat step 3 - 6 several times.
8
Fill syringe again with 20 ml distilled water.
9
Connect the syringe to the filter holder.
10
Empty the syringe with the filter up to the 10 ml mark.
11
Fill the remaining 10 ml filtered sample water into a clean 24 mm cuvette.
12
Screw the lid back on the cuvette.
13
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
14
Start ZERO measurement.
15
Remove the cuvette again.
16
Remove the filter syringe from the filter holder.
17
Fill the syringe with the sample water.
18
Connect the syringe to the filter holder.
19
Empty the syringe with the filter completely.
20
Repeat step 16 - 19 several times.
21
Fill the syringe with the sample water.
22
Empty the syringe with the filter up to the 10 ml mark.
23
Fill the remaining 10 ml filtered sample water into a clean 24 mm cuvette.
24
Screw the lid back on the cuvette.
25
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
26
Tap TEST to perform the measurement.
T60/204
T60/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...