(66)
Hydrogen Peroxide LR
0.00 - 1.90
mg/l (H₂O₂)
Liquid
Internal Name: 66-Hydr-Per-LR-liq
10ml
30ml Hydrogen Peroxide LR N°1 (PL30HydLRP1)
30ml PL Hydrogen Peroxide LR N°2
(PL30HydLRP2)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Empty the cuvette to a few drops.
8
Add 3 drop(s) of PL Hydrogen Peroxide LR 1 into the cuvette.
9
Add 3 drop(s) of PL Hydrogen Peroxide LR 2 into the cuvette.
10
Fill the cuvette to 10 ml with the sample water.
11
Screw the lid back on the cuvette.
12
Gently swirl the cuvette to mix the liquid well.
13
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
14
Tap TEST to perform the measurement.
15
After the lapse of a 02:00 minute(s) countdown the measured value is displayed.
Notes:
If the sample water contains more than 10 mg/l hydrogen peroxide, a measured value of 0 mg/l
can be displayed. In that case, a dilution is recommended.
Make sure no hydrogen peroxide escapes while preparing the sample. The measurement must
be performed directly after sampling.
If the sample water contains further oxidizing agents, these will react like hydrogen peroxide
and contribute to the measurement result.
The DPD reagent buffers the pH value of the sample water in the range of 6.2 to 6.5, which is
essential for colour development. If the sample water is very alkaline or acidic, this must be
adjusted to a pH within the range of 6 to 7 by addition of 0.5 mol/l sulphuric acid or 1 mol/l
caustic soda, respectively, before the DPD reagent is added.
T88/204
T88/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...