Please keep your PrimeLab 2.0 clean!
Do not use detergents to clean your PrimeLab 2.0 but solely use little water and a soft cloth.
Make sure the transparent part (behind the vial-adapter) is properly clean and the vials used are
without fingerprints, dirt, scratches. Always keep the sampling chamber (behind the vial adapter)
clean. On 4 sides of the chamber, you will see small holes behind on a dark plastic part. The LEDs
and sensors are located behind these. All transparent parts in front must be dry and clean.
Avoid water entering the USB port of your PrimeLab 2.0.
The PrimeLab beams light (LED) from one side of the measurement-chamber through the
measurement chamber to the sensor(s) on the opposite or 90° side of the measurement chamber.
Any interference (dirt, fingerprints, scratches) influence the light beam (less transmission) and will
lead to wrong readings / wrong or failed calibration.
Do not exert any pressure when cleaning your PrimeLab 2.0, especially when cleaning the display.
Clean the plastic pane in front of the camera lens to ensure that QR-codes (Available for you soon)
can be recognized properly.
Any soiling must be cleaned properly.
Cleaning the device
PrimeLab
2.0
60
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...