![Tektronix RSA6100A Series Help Manual Download Page 385](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169385.webp)
Reference > Troubleshooting
Error and Information Messages
Increase the Reference Level setting.
This message is just a reminder that the
internal attenuator (see page 307)
has been set to manual mode.
The input signal contains peaks greater than 6 dB above the Reference Level setting.
Increase Reference Level
Decrease input signal power
An error occurred while saving a Setup or Data
fi
le.
Save the
fi
le again.
This is a status message displayed while data is being saved to a
fi
le.
Selected VBW does not use full Spectrum Length
This message can occur when the Spectrum Length is greater than required for the VBW
fi
lter. If you
look at the Time Overview display, the Magenta line for Results Length indicates the part of the Spectrum
Length that was actually used. The measurement results are correct, but don’t include some of the data in
the selected Spectrum Length. To clear this message, you can set the Spectrum Length to Auto.
Setup error: <description of error>
When this message appears, it includes text that explains the problem. For example, the ACPR display
might show: “Setup error: channels can’t overlap”. Setup errors are the result of con
fl
icts in instrument
settings. These types of errors occur when a user makes manual changes to settings. In the example
above, the channel settings in the Channel Power and ACPR display have been set so that the channels
overlap in frequency.
Adjust the instrument settings, or change a setting back to Auto, to eliminate the error.
Setup error: Integration BW exceeds Measurement BW
When this message appears, it includes text that explains the problem. Setup errors are the result of
con
fl
icts in instrument settings. These types of errors occur when a user makes manual changes to settings.
Adjust the instrument settings, or change a setting back to Auto, to eliminate the error.
Setup error: Measurement time for Freq & Phase results
The
Measurement time for Freq & Phase results (see page 214)
speci
fi
es how far across the pulse top the
instrument should wait before measuring the Phase Difference and Frequency Difference for each pulse.
If this value is set too large for any of the pulses in the signal, the measurement point ends up on the
falling edge or during the pulse off time.
Decrease the Measurement time for Freq & Phase results setting (Settings > De
fi
ne tab).
Swept: RF Trig invalid for most signals
RSA6100A Series Real-Time Spectrum Analyzers Help
373
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...