![Tektronix RSA6100A Series Help Manual Download Page 173](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169173.webp)
RF Measurements > Spurious
Reference Tab
Setting the Power Reference Level to Carrier
To set the power reference to Carrier:
1.
Select
Carrier
from the
Power reference
drop-down list. This displays several parameters.
2.
To set the Carrier frequency, click
Edit
. This displays the
Edit Range and Carrier frequencies
dialog box.
3.
Specify the necessary values in the
Frequency
and
Channel width
entry boxes under
Carrier
.
Click
OK
.
4.
Specify the
Threshold
level.
5.
Specify the
Integration BW
. The Integration BW is the bandwidth over which the carrier is measured.
6.
Although Auto is the recommended setting, you can specify the Resolution Bandwidth by changing
the value in the
Res BW
entry box.
7.
Set the
Detection
method. Select either
+Peak
or
Average
.
NOTE.
If the power level measured over the Integration BW about the carrier frequency is below the
threshold level, the instrument concludes that no carrier is present and no relative measurements will be
taken. Absolute measurements will still be taken.
RSA6100A Series Real-Time Spectrum Analyzers Help
161
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...