RF Measurements > Phase Noise
Parameters Tab
Parameters Tab
The Parameters tab enables you to specify acquisition settings that control the Phase Noise measurement.
Setting
Description
Optimization
Selects setting to optimize for speed or dynamic range.
Speed
Optimizes acquisition for speed.
Dynamic Range
Optimizes the acquisition for dynamic range. For this setting, the instrument uses
small bandwidth steps and attenuator optimization to maximize the dynamic range of
the measurement.
Average
Speci
fi
es whether averaging is used or not. This setting affects integrated numeric results
and the trace display. The range for this setting is 2 to 100. The default setting is disabled.
xxx
176
RSA6100A Series Real-Time Spectrum Analyzers Help