Pulsed RF > Common Controls for Pulsed RF Displays
De
fi
ne Tab
Setting
Description
Time method
Speci
fi
es how measurement parameters are determined. The choices are
Absolute and Relative.
Absolute time method
Ref (R)
Ref speci
fi
es the level along the pulse rising edge that is de
fi
ned as the point
from which Start time is counted. Choices: 50%, 60%, 70%, 80%, 90% 100%;
Default: 100%
Start (S)
Start speci
fi
es the time the instrument waits after the rising-edge reference before
starting to measure the pulse. The Start time is measured from the point along the
pulse rising edge speci
fi
ed by the Ref setting. Start Range: ±100 ms; Resolution:
3 digits; inc/dec small: 1 ns, large: 1,2,5,10...; Default: 0
Length (L)
Amount of time that should be included in the measurement. The measurement
time begins at the Start point and continues for the amount of time speci
fi
ed by
Length.
Relative time method
Ref
Ref speci
fi
es the level along the pulse rising edge that is de
fi
ned as the point
from which Start time is counted. Choices: 50%, 60%, 70%, 80%, 90% 100%;
Default: 100%
Start
Start speci
fi
es the time the instrument waits after the rising-edge reference before
starting to measure the pulse. The Start time is measured from the point along the
pulse rising edge speci
fi
ed by the Ref setting. Start Range: ±100 ms; Resolution:
3 digits; inc/dec small: 1 ns, large: 1,2,5,10...; Default: 0
Length
In the Relative time method, Length speci
fi
es the percentage of the top of the pulse
that is used for measurements. The instrument automatically determines the pulse
top. The measurement length is centered within the pulse ON time.
xxx
RSA6100A Series Real-Time Spectrum Analyzers Help
219
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...