Reference > Troubleshooting
Error and Information Messages
An acquisition was completed even though a hardware failure was detected. This message refers to the
acquisition data currently being analyzed, but not necessarily to the current status of the instrument. See
the explanation for "Hardware Failure".
If this data is from a saved
fi
le, the error cannot be cleared.
Data acquired during RF ADC overrange
An acquisition was completed but the signal was outside the range of the analog-to-digital converter at the
time this data was acquired. This message refers to the acquisition data currently being analyzed, but not
necessarily to the current status of the instrument. See the explanation for "RF ADC overrange".
If this data is from a saved
fi
le, this error cannot be cleared.
Data acquired during RF digital gain over
This data was acquired when the input signal contained peaks greater than 6 dB above the Reference
Level setting.
See the de
fi
nition for "RF digital gain over
fl
ow".
If the data is from a
fi
le, this error cannot be cleared.
This data was acquired while the instrument was warming up. The warm-up period is 20 minutes. Until
this warm-up period is completed, the instrument is considered uncalibrated.
Data from unaligned instrument
The acquisition data was captured when the instrument was not aligned. This message refers to the
acquisition data currently being analyzed, but not necessarily to the current status of the instrument.
Measurements made on this data might not be accurate.
Data from uncalibrated instrument
The acquisition data was captured when the instrument was not calibrated. This message refers to the
acquisition data currently being analyzed, but not necessarily to the current status of the instrument.
See the de
fi
nition for "Not calibrated".
If this data is from a
fi
le, the error cannot be cleared.
Data acquired with 3 GHz preamp
This is a warning that signals above 3 GHz are likely to be severely attenuated because the Preamp was on
when the data was acquired. The (optional) Preamp is not speci
fi
ed for operation above 3 GHz.
Disabled: data is from swept acquisition
The display needs to run in real-time mode. The display associated with this message cannot run now
because it is not the selected display, and the selected display is performing multiple acquisitions (it is
in swept mode).
Change the settings of the selected display so it is performing real-time acquisitions.
Select the associated display to make it the selected display. When it is selected, it will force the
acquisition parameters to change to meet its own needs.
368
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...