Analog Modulation > AM
Trace Tab
Average Method
Median Method
In both cases, the instrument computes the average and the median over the analysis period. It should be
noted that the two methods give the same result for sinusoidal modulation where the sampling frequency
is much higher than the modulation frequency.
Trace Tab
The Trace Tab allows you to set the display characteristics of a trace.
Setting
Description
Show
Shows/hides the trace. If the instrument continues to run, the measurement results below
the graph display continue to update even if the trace is hidden.
Detection
Sets the Detection method used for the trace. Available detection methods are +Peak,
-Peak, and Avg (VRMS). Not available for saved traces.
Function
Selects the trace processing method. The only available setting is Normal.
Freeze
Halts updates to the trace.
Save Trace As
Saves the trace to a
fi
le for later recall and analysis.
Show recalled trace
Displays a saved trace instead of a live trace.
xxx
Detection
Detection refers to the method of processing the data acquisition points when creating a trace. The IQ
samples in a data acquisition can be detected in a variety of ways. The number of IQ samples available to
each trace point varies with both analysis length and trace length. For example, with Spectrum Length set
to ‘Auto’ in the Analysis menu, the instrument analyzes just enough samples to produce one IQ sample
pair per trace point. In this case, the detection method chosen has very little effect, as the +Peak, -Peak,
and Avg (VRMS) are all equal. Changing the Spectrum Length causes the available detection methods to
differ in value because they have a larger set of samples for the various detection methods to process.
The available detection methods are:
112
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...