Reference > Menus > View Menu
Status Bar
Status Bar
The Status Bar displays information on speci
fi
c instrument settings. It contains only status information; it
does not display any error information. The Status bar has no controls. It can be hidden.
Elements of the Status Bar
Area within Bar
Description
Run Status
Displays the running state. For example, some run states are Arm, Ready, Triggered,
Analyzing, or Stopped.
Status message area
Displays status messages (for example, ADC Overrange). When there no status
messages to report, displays current acquisition parameters (for example, Acq Length:
22.320 us).
Acquisition mode
Displays either Real Time or Swept. Indicates how the next acquisition will be acquired.
Trigger Settings
Displays Trigger settings. For example, you will see Free Run, Power, Trig In (front),
Frequency Mask and others.
Frequency Reference Setting
Indicates whether the internal or external reference source is selected. Displays Ref:
Int or Ref: Ext.
Internal Attenuator setting
Displays the setting of the internal attenuator.
Preamp Setting
If Option 01 is installed, this indicates whether the preamp is on or off. If the preamp is
on, displays Preamp: ON, 3 GHz max. as a reminder that the input frequency range is
restricted when using the preamp.
xxx
Run Status Indicators
Indicator
Description
Acquiring
The analyzer is capturing the signal.
Analyzing
The analyzer has captured the signal and is processing the signal record.
Arm
The trigger circuitry is
fi
lling the pretrigger portion of the signal record.
File Data
The analyzer is analyzing recalled acquisition records.
Stopped
If Stopped is displayed, Signal acquisition has been halted. This can occur because the
Run button has been pressed or because a trigger event has occurred, signal acquisition
has occurred and Run mode was set to Single.
Triggered
If Triggered is displayed, the analyzer has recognized a valid trigger and is
fi
lling the
posttrigger portion of the signal record.
xxx
Showing or Hiding the Status Bar
Select
View
>
Status
to toggle the display of the Status bar.
RSA6100A Series Real-Time Spectrum Analyzers Help
355
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...