Amplitude Corrections
External Gain/Loss Correction Tab
3.
To edit values in the table:
a.
To edit an existing value, double-click on the cell you want to edit.
b.
To add a new point to the table, click on
Add Point
. Edit the values as required.
c.
Click
Apply
if you want to test your table without closing the editor (the table must be enabled
with the check box in the control panel and measurements must already be running).
To specify a frequency multiplier, you can type K, M, or G.
4.
To delete a point from the table, select the point to be removed, and click
Delete
.
5.
Select the Frequency Interpolation method:
Select Linear when the frequency scale of the spectrum or spurious measurements are linear.
Select Log when the frequency scale of the spectrum is logarithmic.
6.
When you have entered all the necessary values, click
Save As...
(to save the table in a
fi
le) or
click
OK
.
7.
From the Save As dialog, name the
fi
le and save it in the desired location.
External loss tables are saved in CSV (Comma Separated Value) format.
Loading an External Loss Table
To load an external loss table
fi
le:
1.
Click
Edit...
to display the Table Editor.
2.
Click
Load....
to display the Open dialog box.
3.
Navigate to the location of the desired
fi
le, select it and click Open.
4.
Click
OK
.
314
RSA6100A Series Real-Time Spectrum Analyzers Help