General Signal Viewing > RF I & Q Vs Time
RF I & Q vs Time Display
RF I & Q vs Time Display
This is a plot of the baseband In-Phase (I) and Quadrature (Q) components of a modulated carrier. The plot
is in the time domain, with I and/or Q values the Y-axis.
Note that the trace(s) in the RF I & Q vs. Time display can be set to a maximum of 100,000 points
(however, the actual number of trace points can extend up to 1,000,000 points if Max trace points is set
to Never Decimate). If the Analysis Length includes more than the selected Max trace points value, the
trace is decimated (using the method speci
fi
ed with the Detection control) to be equal to or less than the
Max trace points setting (except when Max trace points is set to Never Decimate). This decimated (or
undecimated) trace is what is used for marker measurements and for results export. You can set the Max
trace points on the
Settings
>
Prefs
tab.
To display an RF I & Q vs. Time display:
1.
Select the
Displays
button or select
Setup
>
Displays
. This shows the
Select Displays
dialog box.
2.
From the
Folders
box, select
General Signal Viewing
.
3.
Double-click the
RF I&Q vs. Time
icon in the
Available Displays
box. This adds the RF I & Q vs.
Time icon to the
Selected displays
box.
4.
Click the
OK
button.
Elements of the Display
96
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...