GP Digital Modulation > Trellis Diagram
Trellis Diagram Display
Item
Element
Description
1
Check mark indicator
The check mark indicator in the upper, left-hand corner of the display shows
when the Frequency Deviation vs Time display is the optimized display.
2
Top of Graph
Sets the phase value indicated at the top of the graph. Changing the top value
affects the bottom of graph readout. Also, note that the top of graph setting
interacts with the internal vertical scale setting (which is not user settable)
such that the range between the top and bottom of the graph increases or
decreases automatically.
3
Position
Speci
fi
es the phase value shown at the center of the graph display.
4
Bottom Readout
Displays the value of the phase value shown at the bottom of graph.
5
Autoscale
Adjusts the vertical and horizontal settings to provide the best display.
6
Position
Displays the horizontal position of the trace on the graph display.
7
Scale
Adjusts the span of the graph in symbols.
8
Freq Error
Displays the difference between the maximum and minimum measured
values of the signal frequency during the Measurement Time. The displayed
frequency error is followed by either
Auto
or
Manual
. This indicates the
selected carrier frequency detection method (see Settings > Freq & BW).
xxx
Changing the Trellis Diagram Settings (see page 254)
254
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...