![Tektronix RSA6100A Series Help Manual Download Page 310](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169310.webp)
Analyzing Data > Analysis Settings
Analysis Time Tab
Setting
Description
Analysis Offset
Speci
fi
es the location of the
fi
rst time sample to use in measurements.
Auto
When enabled, causes the instrument to set the Analysis Offset value based on the
requirements of the selected display.
Analysis Length
Speci
fi
es the length of time to use in measurements.
Auto
When enabled, causes the instrument to set the Analysis Length value based on the
requirements of the selected display.
Time Zero Reference
Speci
fi
es the zero point for the analysis time.
Actual
This is a displayed value, not a setting. It is the Analysis Length (time) being used by the
analyzer; this value may not match the Analysis Length requested (in manual mode).
xxx
Analysis Offset
Use analysis offset to specify where measurements begin. Range: 0 to [(end of acquisition) - Analysis
Length)]. Resolution: 1 effective sample (or symbol).
Analysis Length
Use the analysis length to specify how long a period of time is analyzed by a measurement. After you
enter a value, this box changes to show the actual value in use, which is constrained by Acquisition Time.
This setting is not available when Auto is checked. Range: minimum value depends on modulation type
to Acquisition Length. Resolution: 1 effective sample (or symbol).
Time Zero Reference
The analysis offset is measured from this point. Choices are: Acquisition Start or Trigger Point.
Parameter
Description
Acquisition Start
Offset is measured from the point at which acquisition begins.
Trigger Point
Offset is measured from the trigger point.
xxx
298
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...