RF Measurements > Occupied BW & x dB BW
Occupied BW & x dB BW Display
Item
Display element
Description
1
Check mark indicator
The check mark indicator denotes the display for which the acquisition
hardware is optimized. This indictor appears only when the display is the
selected display.
2
Averaging readout
Appears when averaging is enabled (Settings > Parameters tab > Average
results). Displays "Avg: n of m" while averaging the results and "Avg: m"
once the requested number of results have been averaged.
3
Autoscale
Adjusts the vertical and horizontal settings so that the entire trace
fi
ts in the
graph.
4
Trace function readout
Displays either Normal or MaxHold to indicate trace function.
5
Clear
Resets count for Average and MaxHold functions. Enabled only when
Averaging or MaxHold is enabled. Pressing Clear will clear the trace and, if
acquisition is running, restart the averaging or hold process.
6
Graph
Displays the input signal. Shaded areas indicate the measurement bandwidth
(Settings > Parameters tab > Measurement BW).
7
Blue lines
The blue lines indicate where the selected results (selected in the Main
results area) are being measured on the trace.
8
Grid divider
Determines the portion of the display allocated to the graph and detailed
results area. You can move the grid divider all the way to the top or bottom
and any position in between.
9
Detailed results area
Displays additional measurements results.
10
Position and dB/div
Position sets the top of graph value. The dB/div setting is the vertical scale
value.
11
Main results area
Displays results for Occupied Bandwidth (OBW) and x dB bandwidth. Use
the two radio buttons to select which of the two results are illustrated in the
graph with the blue lines and arrows. There are also controls for adjusting
OBW % Power, x dB BW, and RBW.
12
VBW
Enables the VBW (Video Bandwidth)
fi
lter. Displays current VBW
fi
lter setting.
See Setup > Settings >
.
xxx
Detailed Results Readouts
Measurement
Description
Freq Error
The frequency difference between the measured carrier frequency of the signal and the
user-selected center frequency of the analyzer.
OBW Level left
The signal level at the left edge of the occupied bandwidth.
OBW Level right
The signal level at the right edge of the occupied bandwidth.
OBW Ref Power
The total power measured within the measurement bandwidth.
x dB BW Ref Power
The peak power measured within the measurement bandwidth.
xxx
Changing the Occupied Bandwidth Settings (see page 151)
150
RSA6100A Series Real-Time Spectrum Analyzers Help
Summary of Contents for RSA6100A Series
Page 2: ......
Page 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Page 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Page 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...