5. Once transducer is in place, secure with the transducer clamping screw. Do not overtighten. Observe
the upstream/downstream orientation and connect the transducer cable to the computer’s input jack.
If a dual-channel unit, make sure you connect the cable to the input jacks of the correct channel.
Repeat this procedure for the number index transducer.
Installing a 1012T Mounting Track in Direct Mode
Paragraph 3.4 describes the installation procedures that lead up to the automatic selection of transduc-
ers, mounting mode and spacing method. Shown is a Model 1012T mounting track to accommodate
size B3 universal transducers. Examine the figure below which shows a typical Pick/Install Xdcr menu
screen. Note the automatic assignment of model numbers for the transducer and mounting track, plus
the designation of number index.
1. Perform all required menu steps up until the point where the flow computer prompts you to press
<ENT> to finish the transducer install routine. Note the reported number index.
You will be inserting an index pin into this hole and the reference hole. Check to ensure that you have
a matched set of transducers. They both should have the same S/N number but marked with either an
“A” or “B” (e.g., 100A and 100B).
2. The Direct mount configuration uses two track rails, one for each transducer, installed 180
o
apart on
the pipe. One track rail includes the tension screw while the other has a locating stud to support the
chain (see below and on next page).
INSTALLATION – DIRECT MOUNT 180º OPPOSED WITH MOUNTING TRACK
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Direct
Spacing Offset
Minimum
Number Index
4
Spacing Method
Track 1012TP
Ltn Value (in)
0.217
Install Completed?
No
Empty Pipe Set
Channel Not Setup
Zero Flow Adjust
Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Note selection of track
P/N, Mount Mode and
Number Index.
Downstream
Transducer
Chain
Tensioning
Screw
Upstream
Transducer
Track Rails
Transducer
Clamping Screw
Number
Index Pin
TOP VIEW
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......