5-76
CQO:1010PVXFM-3
Section 5
Notice:
All Siemens flowmeters include a real-time clock to provide a convenient date/time stamp for display screens and
datalogger reports. The operating system does not rely on the date and time-of-day for any flowmeter operation.
Therefore, the “Turn Of The Century” has had no effect on the proper functioning of any of our systems. Although we
limit the year to 2 digits in all our displays, the datalogger reports are Y2K complaint using an industry standard
algorithm to form a 4 digit year field.
To perform a 32ºF (0ºC) RTD calibration:
Immerse RTD sensor in deionized water and ice mixture. Stir the mixture constantly.
Move the highlight to [RTD1] by pressing .
To access the [RTD] option list press .
Move the highlight to [User Cal] then press .
This triggers the prompt:
ENT
NOTE: Factory Calibration provides an additional prompt after a new temperature is entered:
[Are you Sure? No Yes]
. It is recommended that you use
[User Cal]
to avoid alteration
of preset factory calibration.
Ice Bath RTD Calibration
Use distilled, deionized water and ice mixture at 32ºF (0ºC) equilibrium for an ice bath. Ensure tempera-
ture with a reference thermometer. Siemens can not assume responsibility for the incorrect design,
construction or operation of an Ice Bath.
CAUTION: DO NOT ALLOW AN RTD SENSOR TO MAKE DIRECT CONTACT WITH ICE
DURING AN ICE BATH CALIBRATION.
When you are sure that the RTD sensor is at 32ºF (0ºC), press to recalibrate the RTD sensor. To
verify the calibrated reading, go to the [Diagnostics/Liquid Data] menu cell to check the current RTD out-
put. Make sure that it coincides with the reading of the reference thermometer.
5.10.8 THE CLOCK SET MENU
The [Clock Set] menu cell allows you to set the time and date. The meter uses its internal clock/calendar
to record the real-time when certain data and diagnostic events occur. In addition, the clock/calendar
provides the Datalogger and Stripchart date and time stamps.
ENT
Calib @ deg F=32.0
Clock Set
Date
xx.xx.xx
(date entry)
Time
xx.xx
(time entry)
CLOCK SET MENU STRUCTURE
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......