5-64
CQO:1010PVXFM-3
Section 5
5.9.3 ANALOG INPUT SETUP (optional function)
The Analog Input Setup function assigns an active analog input to a measurement channel/path. The
meter provides one DC current input port per measurement channel/path. The DC current input ranges
from a zero level of 4 mA to a full scale of 20 mA. The [Analog Input Setup] menu cell allows you to
enable this port and then span it to any desired scaling.
For example, when using the analog input viscosity function the numeric variables might be spanned as
follows: 4mA=1 (water) and 20mA=100 (more viscous liquid). This spanning configuration allows the
meter to use this constant numerical change to improve calibration in real time.
The FUH1010PVX flowmeter models allow you to associate the analog input to active system variables
such as specific gravity, viscosity and others (see tabel below).
NOTE: Refer to the Installation Drawing (1010X-7 and 1010DX-7) or I/O Module markings for the
locations of these inputs and wiring procedures.
I/O Data Control
Analog Inp Setup
Iin1
Input
Off
Aux
cS
PSIA
BARA
T1 Deg F
T1 Deg C
T2 Deg F
T2 Deg C
4 mA
numeric entry
20 mA
numeric entry
Iin2/Iin3/Iin4
See Iin1 option list
Note that the flowmeter recognizes the first analog input variable that is assigned to any given param-
eter and ignores any subsequent input with the same assignment. For example, if Iin1 and Iin2 are both
assigned to represent pressure (PSIA), the meter will only use the pressure input from Iin1.
Setting up the Analog Current Input
The first step is to enable the DC current input port.
From the [Analog Input Setup] menu cell:
To access the [Iin] option list press the <Right Arrow> twice.
Move the cursor down to [Aux (n)] press the <Down Arrow> and then <ENT>. This enables the
port to receive an input current. The cursor moves to [4 mA].
To enable numeric entry press the <Right Arrow>. Type a numeric value corresponding to a 4 mA
input signal.
To assign functions to Relay 1:
To access the [Relay 1] option list press
Move the cursor to the desired relay assignment by pressing
To store selection press
ENT
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......