5-55
CQO:1010PVXFM-3
Section 5
The table may contain up to 32 pairs of these slope correction factors. Note that the
Kc factor, unlike
these slope correction factors
, is entered as a signed percent change in rate, while these factors are
simply rate multipliers. As points are entered, the point editor will provide list access to the already
entered points plus access to the [New Point] menu cell, used to add a new point. The table may be
created in its entirety and then activated by selecting [Yes] in the [Table Active] menu cell. The entire
table may be cleared by selecting [Yes] in the [Clear Table] menu cell.
Note that the meter allows you to use [Aux] as an [Index Variable] input. Effectively you may slope
modulate the meter’s output based on the value of a 4 - 20 mA input signal. Also notable, the interaction
between the Liquid Table, described elsewhere in this manual, and the calibration table(s) is extremely
powerful. Since the table of slope correction coefficients resides between the meter’s internal flow
register(s) and the flow register, which actually provides reports to the user, poorly formed or defective
Calibration Tables may severely impact the performance of the instrument. Take precautions before
enabling these calibration tables. Although we recognize that it is unlikely that all three tables would ever
be employed in a real installation, three tables are offered for maximum user flexibility. Since the tables
can be disabled without being destroyed, 2 or 3 optimization strategies may be tested by this means in
order to determine which approach is most effective.
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......