Special Cautionary Note
Regarding Input Power Wiring
for Compact Explosion Proof Flowmeters
The compact area of the power wiring compartment located under the Rear Housing Cover does not
permit labeling for connecting the flowmeter to its AC or DC power source. The labeled numbers 1,2 and
3 of connector P8 prompt the user to consult the 1010X-7 installation diagram in Appendix B for wiring
details. To avoid damage to the flowmeter, especially in cases where it is difficult or impossible to consult
the 1010X-7 diagram, the correct power supply wiring procedure and corresponding part numbers are
listed below.
1. In addition to confirming the part number of the flowmeter, it is recommended that the actual power
supply module part number be checked for conformation to the power available in the field.
Before
applying power to the unit
, you may visually confirm the part number of the power supply by re-
moving the front housing cover.
2. Once the flowmeter’s power supply type has been established and conforms to the available line
power in the field, the P8 power connector can be wired per the table below.
Root Meter Number
Transducer Support
Relay Support
Numeric Display
Power Supply Option
1010X-T2KNZPL2-S2----Safety Rating Option
Language Option
Typical Part Number Construction:
Power Supply Option Code
Power Supply P/N
User Supplied Power
S
1010X-6SS2
90-230 VAC Single Phase
ZN
010X-6ZNS2
9-36 VDC Negative Ground
ZP
1010X-6ZPS2
9-36 VDC Positive Ground
Terminal
1010X-6SS2
1010X-6ZN
1010X-6ZP
Number
1
Hot
Positive
Positive (Gnd)
2
Neutral
Negative (Gnd)
Negative
3
Ground
Ground
Ground
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......