Section 5
RTD Calibration By Data Entry
The [RTD Calibrate] menu allows you to adjust the intrinsic RTD reading to match an external reference
thermometer by directly entering its reading. Only perform this procedure while the RTD under test is
installed and currently measuring temperature.
To enter the current RTD temperature:
Move the highlight to from [RTD Calibrate] to [RTD 1] by pressing .
To access the [RTD 1] option list press .
Move the highlight to [User Cal] or [Factory] calibration then press . This triggers the prompt:
5.10.7 THE RTD CALIBRATE MENU
(optional)
The RTD Calibrate Menu appears on FUH1010DVX-3 models only.
Use this menu to calibrate the
991T or 1011TN Restive Temperature Device (RTD) temperature sensors to an external standard. It is
important to note that Siemens RTD temperature sensors are factory-calibrated for high accuracy. We
recommend that before deciding to perform the calibration, check the current RTD reading in [Diagnos-
tics/Liquid Data]. You may find that you do not even need to calibrate the sensor. In any case, make sure
that the temperature reading stabilizes before proceeding further. The RTD Calibrate menu allows you to
perform an external calibration, which can be accomplished either by data entry of the current RTD
temperature or by a 32
0
F (0
0
C) Ice-Bath procedure. You can switch between the intrinsic and external
calibration modes at any time.
NOTE: If you perform an external temperature calibration, you should mark and record the
location of each connector and sensor-cable. Once you have re-calibrated the tempera-
ture sensors, changing the sensor/connector orientation established during the proce-
dure may void the calibration.
To access the RTD Calibrate menu press
ENT
To enable numeric entry press , then enter the reading of the reference thermometer (e.g.,
72.0).
ENT
To recalibrate the RTD sensor press . To verify the calibrated reading, go to the [Diagnostics/
Liquid Data] menu cell to check the current RTD output. Make sure that it coincides with the
reading of the reference thermometer.
RTD Calibrate
RTD 1
Factory
User Cal
Calib @ deg F=32.0
Calib @ deg F=72.0
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......