6-5
CQO:1010PVXFM-3
Section 6
6.3
THE APPLICATION INFO MENU
This menu provides a live display of the basic timing data used by the meter during operation.
To activate AnCal:
Move the cursor to the [AnCal] menu cell by pressing . Press the to enable
numeric entry. Note that an equal sign (=) appears before the number.
Enter the desired flow rate using current rate units (e.g., 120.00 GPM). Note that the
[Flow] menu cell now reflects the artificial rate.
Move the cursor away from the menu cell by pressing to turn [AnCal] off.
Flow GAL/MIN
120.00
Total GAL
5645.820
Vs m/s
1274.33
Signal mV
44
Valc %
57
Vaer %
13
Alarm Status
------M--
AnCal GAL/MIN
> 120.000
HiFlow GAL/MIN
1577.9
LoFlow GAL/MIN
-1577.9
Siemens Dual Path
SITE1
Flow Rate Simulation Control
Flow Data
The M status is
asserted due to the
method used to
simulate flow data.
Note that the [Flow]
menu cell immediately
reflects the AnCal entry.
Over time, the Total
display will also
increase.
TN uSec
75.193
TL uSec
44.009
DeltaT
126
Burst/Sec
256
Last Makeup
(Re)Start
Makeup Status
Measurement
Siemens Dual Path
SITE1
Time between transit and receive
Application Info
APPLICATION INFO MENU ITEMS
TN uSec
The total elapsed time between the transmission and reception of a sonic pulse.
TL uSec
The total time a sonic pulse takes to travel through the liquid.
DeltaT nSec
This is the instantaneous result of subtracting the down from the up transit
times.
NOTE: This will appear to be more active than dampened flow readings
.
Burst/Sec
Number of transmissions per second under current operating conditions. The
following factors influence this parameter: aeration, pressure transients, signal
strength variation.
Last Makeup
Reason for last Makeup (signal reacquisition). Press the meter’s [Help] key at
this menu cell to see a glossary of all the possible reasons for a Makeup.
Makeup Status
Current stage of Makeup routine. [Measurement = normal operation]
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......