5-78
CQO:1010PVXFM-3
Section 5
Baud Rate
The [Baud Rate] menu cell sets the asynchronous serial transmission data transfer rate of the RS-232
port. It provides a selection of standard baud rates up to 38,400 baud. The selected baud rate must
match the baud rate setting of the receiving external device. The factory-set baud rate is 9600.
To change the Baud Rate:
To access the [Baud Rate] option list press
Move the cursor to the required baud rate by pressing
To store selection press . This moves the cursor to [Parity].
Parity
Parity is a simple method to check the accuracy of an asynchronous serial data transfer. The parity
setting tells the meter how to format the data words it sends to an external device. Parity is usually an
additional bit added to each data word. For example, if you select [EVEN], the total sum of all the bits in
a single data word (including the parity bit) will always be an even number.
The [Parity] option list includes all the standard parity settings for asynchronous serial transmission. The
selected parity must match the parity setting of the receiving external device. Some devices ignore parity
entirely; therefore, the option list includes [None]. The factory setting is [None].
ENT
RS-232 SETUP MENU STRUCTURE
RS-232 Setup
Baud Rate
300
1200
2400
4800
9600
38400
Parity
None
Odd
Space
Even
Mark
Data Bits
7
8
Line Feed
No
Yes
Network ID
xx
(numeric entry)
RTS Key Time
0.2 Sec.
0.4 Sec.
0.6 Sec.
0.8 Sec.
1.0 Sec.
2.0 Sec.
3.0 Sec.
4.0 Sec.
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......