MOUNTING TRANSDUCERS ON A 1012 TRANSDUCER TEST BLOCK
z
Identify the side of the simulator that applies to the transducers under test. Rotate the clamping
bracket as required to mount transducers on the test surface.
z
Using a coupling compound (preferably CC-102), mount the transducers on the pipe simulator as
shown above. Slide each transducer until it presses against the pin-stop. Use the clamping screws to
hold the transducers in place.
z
Connect transducer cables between each transducer and the meter connectors for the channel
under test. The Up and Down orientation is not important.
z
Access the Installation Menu. Select [Meter Type] [Single, Dual, or Quad Channel] depending on
meter type. Select the meter channel (1,2,3, or 4) depending on which measurement channel you
intend to test. Select [Clamp-On] and then [Channel Setup].
z
Access the [Channel/Path Setup] menu. Move the highlight to [Create/Name Site]. Create a new Site
Setup (e.g., TEST1). You can now enter data without altering an existing Site Setup.
z
Select the [Pipe Data] Menu. Referencing either the English or metric pipe simulator
chart below, enter the pipe data corresponding to the transducer size under test.
Transducer clamping
screw
Transducer testing
surface for an "A"
size transducer.
Slide transducers firmly against
transducer pinstops.
Transducer
under test
Transducer
under test
Transducer
Clamping
Bracket
This example shows testing
for Size B. Rotate brackets
180 degrees to mount Size
A transducers.
This label shows the
data entry para-
meters for Size B.
Size A parameters
appear on a label
on the other side
plate.
This arrow points to the surface
associated with the Xdcr size
identified on the label.
6.9.1 USING THE 1012TB-1 AND 2 TEST BLOCKS
The 1012TB-1 and 1012TB-2 test blocks provide two test surfaces. Each surface supports a specific
transducer size. For example, one surface of the 1010TB-1 supports Size “A” transducers and the other
supports Size “B” transducers. The 1012 pipe simulators include two labels, one on each side-plate. The
labels identify the transducer size, data to be entered, and the surface to be used with the specific
transducer size. See drawing below.
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......