This completes the ReversaMatic procedure. The system’s zero accuracy will be very close to that obtain-
able using the Actual Zero method, providing flow remained constant during this procedure.
Important Note:
A caution on the use of upper and lower flow limits (used to prevent flow mis-
registration) prior to using the Reversal Zero technique (ReversaMatic): If the negative flow rate
that the meter reads in the step during which the transducers are reversed is more negative than
the lower flow limit, the meter will re-register positive and the Reversal Zero cycle will thus be
corrupted.
Therefore, postpone the installation of upper and lower flow limits until the reversal zero proce-
dure is executed successfully. For pipes that combine large diameters with very high flow veloci-
ties, it may be necessary to move the upper and lower flow limits out of the way until the reversal
zero is completed. Moreover, pipes of this size frequently have excellent intrinsic zero perfor-
mance and may not even need zeroing.
ZeroMatic
(optional function)
REYNOLDS TABLE
NOTE: ZeroMatic is used in the Reflect Mode only.
Use this menu cell to select the ZeroMatic option. If conditions permit the use of the Auto Zero function
then the ZeroMatic option can be used as well. ZeroMatic is similar to Auto Zero, but differs in that its
averaging function is continuously updated and virtually eliminates zero drift.
To select and enable the ZeroMatic option:
In the Pick/Install Xdcr menu, press to scroll to the [Zero Flow Adjust] menu cell.
To access the [Zero Flow Adjust] option list press
NOTE: If ZeroMatic is not running, the [Actual Zero] menu item will be displayed next to the
[Zero Flow Adjust] menu cell.
Siemens Dual Path
SITE1
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Reflect
Spacing Offset
Minimum
Flow
Number Index
7
Spacing Method
Track 1012TP
Ltn Value (in)
ReversaMatic
Install Complete?
ZeroMatic
Empty Pipe Set
Auto Zero
Zero Flow Adjust
Actual Zero
ZeroClr
Conformans Indicated flow to Actual Zero
Pick/Install Xdcr
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......