5-19
CQO:1010PVXFM-3
Section 5
Each of the data points of a Site Liquid Table (maximum of 32) may be filled in by the user as indicated
in the menu screen shown above.
Index Value -
The temperature-corrected sonic velocity (LiquIdent) which points to the output variables
forming the balance of the table entry. Below this you may enter the values associated with that LiquIdent
value.
S.G.
- The specific gravity of the liquid at system Base Temperature, usually 60ºF. By specific gravity we
mean the ratio of equal volumes of this liquid and water at this base temperature.
Viscosity
- The kinematic viscosity of the liquid at system Base Temperature (units: centistokes).
Visc Slope
- The exponent of the logarithmic expression used to project the liquid viscosity at measured
temperature. The default value (-0.0287) has been found to be adequate for many hydrocarbons.
γγγγγ
n
=
e
where:
γ
n
= Kinematic viscosity at current measured temperature (cS)
e
= Base of natural logarithms
V
SLP
= Viscosity Slope (i.e., -0.0287)
Ì
t
= Current Measured Temperature minus Base Temperture (degrees F)
γ
b
= Visc @ Base Temperature (cS)
(
)
V
SLP
x
Ì
Ì
Ì
Ì
Ì
t + ln(
γγγγγ
b
)
Siemens
Dual Path
Site 1
LiquIdent Index
Index Value
1440
Visc (cS) @ 400
1.00
Visc Slope @ 400
-0.0287
Kill Index @ 400
No
Temperature Corrected Vs Index
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......