5-53
CQO:1010PVXFM-3
Section 5
Intrinsic
When selected, the meter uses no slope adjustment at all. Output data is still zeroed and corrected for
Reynolds number, but no slope adjustment is imposed on the meter’s flow register.
Some applications may require an output adjustment to match an official external reference. The [Cali-
brate Flowrate] menu allows you to select a calibration mode. The right-hand column shows the active
calibration mode. You can select Intrinsic (factory) and Kc (Slope Correction) Calibration. Selecting either
of the external calibration modes will not eliminate the Intrinsic (factory) calibration. You can use this
menu cell to switch between Intrinsic and Kc at any time.
Kc Calibration
For most applications, the measured flow range produces a linear meter response. Therefore, the Kc
(slope correction) calibration is the preferred method since it only requires a single correction factor for
all the flow rates encountered.
NOTE: Changing the calibration can cause profound changes in a flowmeter’s operating charac-
teristics. Use only the most respected flow standard to obtain a correction factor. The
percentage you enter must provide an accurate and consistent shift across the entire
flow range anticipated for the application.
Data Span/Set/Cal
Calibrate Flowrate
Intrinsic
Kc
x.xxx
Calib. Table 1
Index Variable
Flow
Vs
Liqident
Viscosity (cS)
Viscosity(cP)
API
S.G.
Valc
Vaer
Temp1
Reynolds #
Pressure
Aux
Calib. Table 1
New Point
Cal. Index Value
x.xxx
PosFlow Corr
x.xxx
NegFlow Corr
x.xxx
Accept/Clear
Accept Pt.
Clear Pt.
Table Active 1
No
Clear Table 1
No
Yes
Data Span/Set/Cal Menu (Calibration Section)
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......