3-1
Section 3
CQO:1010PVXFM-3
3.
HARDWARE INSTALLATION GUIDE
This section explains how to mount the flow display computer assembly and how to complete the Input/
Output wiring. In addition, it provides details on how to select a location and then install clamp-on trans-
ducers. Please refer to the engineering drawings located in Appendix B that provide important installation
information that does not appear in this section.
3.1
MOUNTING THE FUH1010DVX FLOW COMPUTER
The FUH1010PVX Flow Display Computer is shown below in the standard upright mounting position.
Please refer to Engineering Drawing 1010X-8 in Appendix B of this field manual for outline dimensions,
etc.
NOTE: All enclosure lids are secured with setscrews. This is required to maintain the explosion-
proof ratings. You will need a 1/16" hex key to remove and replace enclosure lids.
STANDARD UPRIGHT MOUNTING
Front Housing Cover
Set screws for
enclosure lid (1/16” hex)
Rear Housing
Cover
Magnetic Wand
To “Downstream” Transducer
To “Upstream” Transducer
1012CNF Series Cable
Customer’s Conduit Fitting
or Cable Gland
Cover Set Screw
Optional Local Temperature Sensor
or 1012EC Series Cable to Remote
Sensor
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......