6-8
CQO:1010PVXFM-3
Section 6
The limits of the “HF” parameter are +/- 0.7 and any attempt to manually install a larger value
will cause the flowmeter to abort the installation of the parameter.
NOTE: Pressing the <Left Arrow> at any stage prior to accepting the measured value will abort
the installation and return to the previous setting.
To access this [HF] menu item proceed as follows:
At the [Meter Type] Menu, press the <Right Arrow> and then <ENT> to select the desired Path
(e.g., Dual Path Flow).
In the [Dual Path Flow] Menu, press the <Down Arrow> and scroll to the [Diagnostic Data] menu
cell. Press the <Right Arrow> to select it.
In the [Diagnostic Data] Menu, highlight [Path Select] and select the desired transducer path.
Press <ENT> to select path.
Press the <Down Arrow> and scroll to the [Site Setup Data] menu cell. Press the <Right Arrow>
to select it.
Siemens
Dual Path
4SS10G
Transducer Setup Data
Path Select
1
Path Enable
Yes
Flow Rate
Application Info
Liquid Data
Site Seup Data
Test Facilities
Print Site Setup
No
Site Created:
04.09.0314.52.23
Diagnostic Data
“Manual” Adjustment Procedure
In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press
the <Right Arrow> and a pop-up [Manual] prompt will appear as shown below.
Note: Press the
<Up/Down Arrow> to select [Automatic], if desired.
Use the numerical keys to input the desired correction value. Press <ENT> to input value.
Siemens
Dual Path
4SS10G
Access Path/Channel Diagnostic Data
Channel /Path Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
Data Span/Set/Cal
Datalogger Setup
I/O Data Control
Diagnostic Data
Dual Path Flow
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......