3-6
CQO:1010PVXFM-3
Section 3
We recommend Reflect mount whenever possible. This is the simplest way to mount the transducers.
Also, Reflect mount resists abnormal flow profile conditions such as cross-flow within the flow stream.
Reflect mode supports the AutoZero function, which zeroes the meter automatically without user-partici-
pation. In addition, Reflect mount may be the only possibility if conditions do not allow access to the
opposite side of the pipe.
Direct mount provides a shorter sonic beam path. This usually improves performance with sonically
attenuative liquids or pipe materials. We recommend using Direct mount for plastic pipes. Compared to
Direct mounting, Reflect mount requires almost double the amount of a straight pipe run. Therefore,
Direct mount may be the only option if the availability of mounting space is limited.
3.3.4 PREPARING THE PIPE
z
Pick a mounting location with the longest straight run. You must have easy access to at least one side
of your pipe. The mounting location must remain full, even at zero flow.
z
Decide on your mounting mode (Direct or Reflect). Always use Reflect Mode whenever possible. You
may only need to use Direct Mode if your pipe is plastic.
z
After receiving the spacing dimensions from the Installation Menu, prepare the pipe surface. De-
grease the surface, if necessary, and remove any grit, corrosion, rust, loose paint, etc. Use abrasive
material provided to provide a clean contact surface for the transducers.
z
Refer to the next sections for illustrated instructions on how to locate each area to be cleaned and
how to use each mounting option.
Please note that the instructions show vertical mounting for clarity purposes only. Do
not install transducers on the top of a pipe.
DIRECT MOUNT
(Pipe shown in top view
for ease in visualizing sonic path)
1234567890
1234567890
1234567890
1234567890
123456789
123456789
123456789
123456789
Cleaned areas. Clean an area 1/2” on either side of
transducer and half again the length of the trans-
ducer on the front and back.
Pipe
PREPARING THE PIPE SURFACE
Upstream
Transducer
Sonic Path
Pipe
Downstream
Transducer
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......