3-15
Section 3
CQO:1010PVXFM-3
3.3.9 USING 1012T MOUNTING TRACKS
The 1012TP and 1012THP Mounting Tracks provide a rigid mounting platform for Series 1011 universal
or high precision size A or B transducers. The mounting tracks service pipe sizes up to a maximum of
5.00” (140 mm) outer diameter. Operating temperatures are supported up to 250°F (121°C). The as-
sembly consists of lightweight aluminum track rails with integral transducer clamping screws. Attached
index pins enable positive locating of the transducers at fixed spacing locations. Roller-chains and ten-
sion screws secure the assembly to the pipe. The following instructions refer to 1011 universal trans-
ducers. Please refer to Engineering Drawing 1012THP-7A for reflect mounting of high precision trans-
ducers and 1012THP-7B for direct mounting of high precision transducers.
The 1012T mounting tracks support both Direct and Reflect mounting modes. The flow computer recom-
mends the appropriate transducers, mounting track and mounting mode, based on the pipe data en-
tries. Refer to the instructions in paragraph 6.4 for details on the Transducer Installation procedure. If
necessary, review paragraphs 3.3.2 through 3.3.4 for details on how to select and prepare a mounting
location on your pipe.
Installing a 1012T Mounting Track in Reflect Mode
Paragraph 3.4 describes the Transducer Installation procedures that lead up to the automatic selection of
transducers, mounting mode and spacing method. Examine the figure below, which shows a typical
Pick/Install Xdcr menu screen. Note the automatic assignment of model numbers for the transducer and
mounting track, plus the designation of the number index
.
This example requires a Model 1012T Mounting Track to accommodate size B3 universal transducers.
Note the reported number index. You will be inserting an index pin into this hole on the track rail to
position one of the transducers (see diagram on next page).
1. Perform all required menu steps up until the point where you respond to the [Install Completed?]
prompt. Note the reported number index. You will be inserting index pins into this hole and the
reference hole on the track rail (see diagram on next page). Check to ensure that you have a matched
set of transducers. They both should have the same S/N number but marked with either an “A” or “B”
(e.g., 100A and 100B).
Note selection of
type, size and
mounting mode.
Note selection of
mounting track part
number and number
index.
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Reflect
Spacing Offset
Minimum
Number Index
6
Spacing Method
Track 1012TP
Ltn Value (in)
0.581
Install Completed?
No
Empty Pipe Set
Channel Not Setup
Zero Flow Adjust
Channel Not Setup
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......