5-58
CQO:1010PVXFM-3
Section 5
Alarm Letter Codes and Descriptions
S ................. Spacing .............. Transducer spacing may need re-adjustment
0 ................. ZeroMatic ........... ZeroMatic signal fault
E ................. Empty ................. Pipe is empty
R ................. Rate ................... Flow above High setting or below Low setting
F ................. Fault ................... Three continuous seconds without new data update
A ................. Aeration .............. Current aeration percentage exceeds the alarm setpoint
M................. Memory .............. Last valid reading for a selected interval during Fault condition
K ................. Makeup .............. In-Process Makeup occurred
I................... Interface ............. Liquid Vs exceeds interface alarm setpoint
P.................. Pig ...................... Pig passage detected (optional)
NOTE: The time interval-based Datalogger records the state of all the alarms at the selected
report time only. Therefore, if an alarm condition resets before the report time, it would
not appear on the report. Use Datalogger Events to log transient alarms. This mode
generates a report upon the detection of an alarm event.
If you select Alarms, the Datalogger logs the state of each alarm function upon generating a report. A
dash [-] represents an inactive alarm. A letter code represents an active alarm (see next page).
A letter code represents an active alarm as shown below.
None
Default – Datalogging is disabled.
Site ID
You entered this site name when you created the site.
Date
The report date according to the meter’s real time clock. (Format: MM.DD.YYYY)
Time
The report time according to the meter’s real time clock. (Format: HH.MM)*
Path Flow
The instantaneous flow measured at the time of the report (for Dual Path Flow meters).
Flow
The instantaneous flow measured at the time of the report.
Average Flow
The average flow rate measured since the last report.
Raw Flow
The uncompensated flow rate in in/sec2 (only zeroed - no Reynolds number curve applied).
Total
The total flow accumulation measured at the time of the report.
Path Vs
The average liquid sonic velocity since the last report.
Vs
The average liquid sonic velocity since the last report.
Path Valc
The signal strength measured at the time of the report (for Dual Path Flow meters).
Valc
The signal strength measured at the time of the report.
Aeration
The aeration percentage recorded at the time of the report.
LiquIdent
Select LiquIdent function.
Path Alarms
The letter codes of any Path Alarms active at the time of the report (see next page).
Alarms
The letter codes of any alarms active at the time of the report (see next page).
Temperature
The instantaneous supply and return temperatures at the time for the report.
Path Delta T
The Up to Down transit-time difference measured at the time of the report (in u/s).
Analog Inputs
Any data appearing on an auxiliary input at the time of the report.
*The Time field expands to HH.MM.SS when a logging interval of more than once per minute is selected.
Summary of Contents for SITRANS FUH1010PVDX
Page 2: ......
Page 4: ......
Page 8: ......
Page 12: ......
Page 20: ......
Page 24: ......
Page 32: ......
Page 184: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 216: ......
Page 219: ...21614 C 1010X 7 CAUTION ...
Page 222: ...STANDARD UPRIGHT INSTALLATION 21614 C 1010X 8 ...
Page 223: ...21614 C 1010DX 7 CAUTION ...
Page 227: ...21614 C 1010DX 8 STANDARD UPRIGHT INSTALLATION ...
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ...1011HNFS 7 21614 C ...
Page 235: ...1011HNFS 7 C 21614 ...
Page 237: ......
Page 241: ...1011NFPS 7 C 21614 REFLECT MODE OPERATION DIRECT MODE OPERATION AUXILIARY VIEW ...
Page 242: ...1011NFPS 8 21614 C ...
Page 243: ......
Page 249: ......
Page 250: ......
Page 251: ...1012TB 8 INSTALLATION OUTLINE 1011 SERIES TEST BLOCK 21614 C ...
Page 254: ...1012TN 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TN SERIES 21614 C MOUNTING TRACK ...
Page 257: ...1012TNH 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TNH SERIES 21614 C MOUNTING TRACK ...
Page 258: ...1012F DB 7 21614 C ...
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ...21614 C OUTLINE DIMENSIONS WELD SEAL TYPE 1013WS 8 ENCLOSURES ...
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 269: ...21614 C TEMPERATURE SENSOR DEDICATED NEMA 4 OUTLINE DIMENSIONS 991TN 8 5 4 3 2 1 ...
Page 270: ......
Page 271: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 286: ......
Page 304: ......
Page 307: ......