P051 Maximum Volume [6 relay model]
For Readings in volumetric units (rather than percent), enter the equivalent vessel volume
for Span (P007).
Any volume units can be chosen because volume is calculated from empty to maximum
span and is scaled according to the Tank Shape (P050) value.
Values
8
Sphere
9
Universal Linear
10
Universal Curved
Alters
•
P001 Operation
•
P051 Maximum Volume
•
Pump Efficiency (P180-P186)
•
Pumped Volume Totalizer (P622-P623)
•
P920 Reading Measurement
Primary Index
Level
Values
Range: 0.000 to 99999
Preset: 100.0
Alters
P060 Decimal Position
Related
•
P006 Empty
•
P007 Span
•
P924 Volume Measurement
Note:
Make sure selected chosen units allow LCD volume display.
Examples:
•
If max. volume = 3650 m
3
, enter 3650
•
If max. volume = 267500 gallons, enter 267.5 (thousands of gallons)