Page 172
HydroRanger 200 – INSTRUCTION MANUAL
7ML19985FC03
mmm
mm
Pa
ra
m
e
te
rs
P610 Head Breakpoints [6 relay model]
The head breakpoints for which flowrate is known. See “Accessing a Secondary Index”
on page 115 for more information..
The values in the Span for which flowrates are known. See
Universal Calculation Support
on page 81 for how to specify universal flows.
P611 Breakpoint Flowrates [6 relay model]
The flowrate corresponding to each Head Breakpoint entered.
These are the flowrates for the related breakpoints. See
Universal Calculation Support
on
page 81 for how to specify universal flows.
Head vs. Flowrate (P610 and P611)
Primary Index
Single Point Model
Dual Point Model
Global
Transducer
Secondary Index
Breakpoint
Values
Range: 0.000 to 9999
Related
•
P611 Breakpoint Flowrates
Primary Index
Single Point Model
Dual Point Model
Global
Transducer
Secondary Index
Breakpoint
Values
Range: 0.000 to 9999
Related
•
P610 Head Breakpoints
Max. Flow
Zero Flow
Ze
ro
Hea
d
Ma
x. Head
Head Breakpoints
Fl
ow
Br
ea
kp
oi
nts