P275 Scaled Discrete Input Value
Shows the current value of the discrete input after any scaling is applied.
Readings are updated continuously even in PROGRAM mode. The value signals a level
override event.
Standard Data Logging (P300 to P321)
All records can be reset by pressing the CLEAR
keys.
Record Temperatures (P300 to P303)
These features display the high and/or low temperatures in °C. When a parameter
relating to a TS-3 Temperature Sensor is accessed, the Point Type display changes to the
TS-3 symbol .
If the unit is powered up without a temperature sensor connected, the value –50°C is
displayed. This information can help trace problems with both built in and external
temperature sensors.
P300 Temperature, Transducer Maximum
Shows the highest temperature encountered, as measured by the temperature sensor in
the transducer (if applicable).
Press CLEAR
keys to reset the log after a short circuit on the transducer wiring.
Primary Index
Discrete input
Values
Display: view only
Values: dependent on the function of the discrete input
Range of Values
Function (P270)
1
Forced
ON
0
Forced
OFF
0 (DI open), 1 (DI closed)
Normally Open
0 (DI closed), 1 (DI open)
Normally Closed
Primary Index
Transducer
Values
Range: - 50 to +150°C (view only)
Preset: - 50°C
Related
•
P301 Temperature, Transducer Min