HydroRanger 200 – INSTRUCTION MANUAL
7ML19985FC03
mmm
mm
Pa
ra
m
e
te
rs
P215 20 mA Output Trim
Calibrates the 20 mA output.
Adjust this value so the device indicates 20.00 mA when P215 is accessed.
mA Output Failsafe (P219) [6 relay model]
Please note that this parameter applies to the HydroRanger 200 6 relay model only.
P219 mA Output Failsafe [6 relay model]
Use for failsafe operation, independent of the Failsafe Material Level (P071).
Selecting an independent mA Failsafe option:
1.
Press MODE
to display the Auxiliary Function symbol.
2.
Press ARROWS
to scroll access the failsafe options.
3.
Press ENTER
when the desired option displayed.
Or, to produce an mA output at a specific value, enter the value required. This is used only
if mA output is allocated to a transducer (P201 = 1 to 7).
mA Input (P250 to P260) [6 relay model]
Please note that these parameters apply to the HydroRanger 200 6 relay model only.
Primary Index
mA output
Values
Preset: 20.00
Related
•
P214 4mA Output Trim
Primary Index
mA output
Values
Range: 0.000 to 22.00
OFF
*
mA output responds to Failsafe Material Level (P071).
HOLd
last
known
value is held until normal operation
resumes
LO
produce the
Empty
mA output immediately
HI
produce the
Span
mA output immediately
Related
•
P201 mA Output Function